S8000 Integrale
High Performance Precision Dewpointmeter
The S8000 Integrale continues Michell's tradition of excellence in the design of high performance humidity products. Building from the solid foundations of performance and reliability in its predecessor, the S4000, this new generation dewpointmeter adds a multitude of new features and user benefits.
The S8000 Integrale offers a wider measurement range, major improvement in response speed, a frost assurance system that eliminates any ambiguity over ice/super-cooled water formation, easy mirror cleaning, multi-language display, vertical and horizontal orientations amongst many other improvements.
Product Features
- Fundamental, accurate and drift-free measurement
- 1 mK display resolution
- -60 to +40°C dew point range with 0.1°C accuracy
- FAST - guaranteed frost formation below 0°C
- Data logging to USB or SD card
- Multi-language display
- Vertical or horizontal configuration

Technical Specification
| GENERAL | |
|---|---|
| Measurement Range | -60°C to +40°C dew point |
| Measurement Accuracy | ±0.1°C dew point |
| Stability | <0.05°C |
| Repeatability | <0.05°C |
| Measurement Units | °C, °F dew point; % rh; °C, °F temperature; gm-³, gkg-¹, ppmv; ppmw; SF6; mlmin-¹ |
| Response Speed | 2°C sec-¹ + settling time |
| Power Supply | 85 to 264 V, 47/63 Hz; 100 VA |
| Operating Temperature Range | -20 to +50°C |
|
DEW-POINT SENSOR |
|
| Mirror | Gold plated copper |
| Temperature Measurement | 4-wire 100 Ω PRT 1/10 DIN class B |
| Pressure | 2 MPa (max) |
| FLOW SENSOR | |
| Measurement Range |
0-1000 mlmin-¹ |
| DISPLAY UNIT | |
| Resolution | User-selectable to 0.001 dependant on parameter |
| Outputs | Analogue: Three channels; user-selectable 4-20 mA, 0-20 mA or 0-1 V Digital: USB, SD Card slot Alarm: Two volt free changeover contacts; 1A @ 30V dc |
| Cable Pack | Supply and USB cables. Output connectors suitable for indoor use |
| OPTIONAL REMOTE PRT | |
| Temeprature Measurement | 4 wire 100 W PRT 1/10 DIN class B |
| Measurement Accuracy* | ±0.1°C |
| Measurement Units | °C or °F |
| Cable Units | 2 metres (250 metres max) |
| OPTIONAL PRESSURE SENSOR (BUILT-IN) | |
| Measurement Range | 0 to 20 Bara |
| Measurement Accuracy* | 0.25 % Full Scale, typical |
| Measurement Units | Bara; Psia; Kpa |
Options:
Microscope
Rack-mount kit (horizontal version)
UKAS Calibration
Transportation case
*Measurement accuracy means maximum deviation between instrument under test and corrected reference. To this must be added the uncertainties associated with the calibration system and the environmental conditions during testing or subsequent use.